PURELAB Classic UV UF (US only)

$0.00
SKU
CLXXUVFM2-US

Description

PURELAB Classic UV UF (US only)

Conductivity

<30 µS/cm (RO)

Temperature

10-15 °C

System Voltage

24V DC

Part Number

CLXXUVFM2-US

Silica

<2 ppm

Organics (TOC)

<3 ppb

Particles

Ultra Filtration

Power Consumption

60 VA (during recirculation)

Footnotes

Bacteria - Specification subject to use with a POU (point of use) filter. If POU filter not used <1 CFU/ml

Reservoir Connection

Jack plug 3.5 mm

Resistivity

1 MΩ-cm minimum at exhaustion (SDI)

Breite (mm)

41016

Bacteria

<0.1 CFU/ml (<1 CFU/10ml)

Brand Name Range

PURELAB

Carbon Dioxide

<30 ppm

Daily Usage (8hr day)

< 100 l/day

Applications

AAS (Atomic absorption spectrophotometry), Bacterial cell culture, Buffer and media preparation, Clinical biochemistry, Electrochemistry, Electrophoresis, Electrophysiology, ELISA (Enzyme linked immunoabsorbant assay), Endotoxin analysis, Flame-AAS, GC-MS, General chemistry, GF-AAS (Graphite Furnace - Atomic Absorption Spectrophotometry), Histology, HPLC, Hydroponics, IC (Ion chromatography), ICP-AES (Inductively Coupled Plasma - Atomic Emission Spectrometry), ICP-MS (Inductively Coupled Plasma - Mass Spectrometry), Immunocytochemistry, Mammalian and bacterial cell culture, Media preparation, Microbiological analysis, Molecular biology, Monoclonal antibody research, Plant growth cabinets, Plant tissue culture, Qualitative analyses, RIA (Radioimmunoassay), Sample dilution and reagent preparation, Solid phase extraction, Spectrophotometry, Stability chambers, TOC analysis, Trace metal detection, Ultra sonic bath, Ultra trace analysis, Water analysis

Depth (mm(inch))

365 (14.4)

Höhe (mm)

49019

Delivery Flow Rate

up to 2 l/min

Footnotes

Bacteria - Subject to use with a POU (point of use) filter

DNase

Removed

Inorganics

18.2 MΩ-cm

Installation Options

Bench, Wall

Mains Input

100-240V ac, 50-60 Hz

Max. Pressure

0.7 bar (10 psi)

Min. Pressure

0.07 bar (1 psi)

Noise Level

<40 dBA

Fuses

2 x T6.3 amp

Drain (gravity fall with air gap)

<2 l/min

Endotoxins

<0.001 EU/ml

Footnotes

Vstupní voda upravená demineralizací může mít vliv na vyčerpání kationty

Feedwater Nature

Pre-treated (RO/SDI/Distilled)

Flow Rate

<130 l/hr

Fouling Index

<1

Free Chlorine

<0.05 ppm

Ships within

4 weeks

Betriebs-Gewicht (kg)

1533

Shipping Depth

19

© 2022 Veolia Water Technologies. All Rights Reserved.